초록

We present an energy-efficient reception and transmission scheme based on the exploitation of the conductive properties of the body for wireless neural interface applications. High efficiencies were achieved with: 1) a three-level clock-embedded direct digital signaling scheme in the transmitter (TX) and 2) with a squaring clock extraction enabling data recovery scheme using a clocked-comparator in the receiver (RX). Additionally, the output voltage of the TX was boosted to secure the link budget of the RX. As a result, the RX and TX consumed 62 and 566 mu W, respectively, for a rated output power of 185 mu W and achieved a maximum data rate of 16.7 Mb/s. These operational characteristics corresponded to the energy consumptions of 3.7 pJ per received bit and 34 pJ per transmitted bit, demonstrating a promising potential for use in implanted devices.

키워드

Body channel communication (BCC); closed-loop neural interface; implantable communication; low-energy communication; neural implant; TRANSCEIVER
제목
An Implantable Body Channel Communication System With 3.7-pJ/b Reception and 34-pJ/b Transmission Efficiencies
저자
Yuk, Beomjin; Kim, Byeongseol; Park, Sanggeon; Huh, Yeowool; Bae, Joonsung
DOI
10.1109/LSSC.2020.2978852
발행일
2020
유형
Article
저널명
IEEE SOLID-STATE CIRCUITS LETTERS
권
3
페이지
50 ~ 53