An Implantable Body Channel Communication System With 3.7-pJ/b Reception and 34-pJ/b Transmission Efficiencies
- Yuk, Beomjin;
- Kim, Byeongseol;
- Park, Sanggeon;
- Huh, Yeowool;
- Bae, Joonsung
WEB OF SCIENCE
11SCOPUS
20초록
We present an energy-efficient reception and transmission scheme based on the exploitation of the conductive properties of the body for wireless neural interface applications. High efficiencies were achieved with: 1) a three-level clock-embedded direct digital signaling scheme in the transmitter (TX) and 2) with a squaring clock extraction enabling data recovery scheme using a clocked-comparator in the receiver (RX). Additionally, the output voltage of the TX was boosted to secure the link budget of the RX. As a result, the RX and TX consumed 62 and 566 mu W, respectively, for a rated output power of 185 mu W and achieved a maximum data rate of 16.7 Mb/s. These operational characteristics corresponded to the energy consumptions of 3.7 pJ per received bit and 34 pJ per transmitted bit, demonstrating a promising potential for use in implanted devices.
키워드
- 제목
- An Implantable Body Channel Communication System With 3.7-pJ/b Reception and 34-pJ/b Transmission Efficiencies
- 저자
- Yuk, Beomjin; Kim, Byeongseol; Park, Sanggeon; Huh, Yeowool; Bae, Joonsung
- 발행일
- 2020
- 유형
- Article
- 저널명
- IEEE SOLID-STATE CIRCUITS LETTERS
- 권
- 3
- 페이지
- 50 ~ 53
- 언어
- ENG
- 출판사
- IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
- 발행국가
- 미국
- 분량
- 4 페이지
- ISSN
- E 2573-9603
P 2573-9603