Computational Quantification of the Cardiac Energy Consumption during Intra-Aortic Balloon Pumping Using a Cardiac Electromechanics Model

Citations

WEB OF SCIENCE

6
Citations

SCOPUS

8

초록

To quantify the reduction in workload during intra-aortic balloon pump ( IABP) therapy, indirect parameters are used, such as the mean arterial pressure during diastole, product of heart rate and peak systolic pressure, and pressure-volume area. Therefore, we investigated the cardiac energy consumption during IABP therapy using a cardiac electromechanics model. We incorporated an IABP function into a previously developed electromechanical model of the ventricle with a lumped model of the circulatory system and investigated the cardiac energy consumption at different IABP inflation volumes. When the IABP was used at inflation level 5, the cardiac output and stroke volume increased 11%, the ejection fraction increased 21%, the stroke work decreased 1%, the mean arterial pressure increased 10%, and the ATP consumption decreased 12%. These results show that although the ATP consumption is decreased significantly, stroke work is decreased only slightly, which indicates that the IABP helps the failed ventricle to pump blood efficiently.

키워드

Intra-Aortic Balloon PumpCardiac Electromechanics ModelATP ConsumptionStroke WorkHEART-FAILURE
제목
Computational Quantification of the Cardiac Energy Consumption during Intra-Aortic Balloon Pumping Using a Cardiac Electromechanics Model
저자
Lim, Ki MooLee, Jeong SangGyeong, Min-SooChoi, Jae-SungChoi, Seong WookShim, Eun Bo
DOI
10.3346/jkms.2013.28.1.93
발행일
2013-01
유형
Article
저널명
Journal of Korean Medical Science
28
1
페이지
93 ~ 99