Environmental gamma-ray dose rate measurement by using ultra-high sensitive LiF:Mg,Cu,Si TLD

  • Kim, M. J.
  • Kim, S. Y.
  • Lee, J. I.
  • Kim, J. L.
  • Chang, I.
  • ... Hong, D. G.
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

5

초록

Environmental gamma-ray dose rates were measured by using ultra-high sensitive LiF:Mg,Cu,Si thermoluminescence dosimeters (TLDs) during short-term burial. The resultant gamma-ray dose rates were compared with those measured by using optically stimulated luminescence (OSL) of Al2O3:C and assessed by using HPGe spectrometer. Regardless of the short-term burial and the change of environmental conditions as well as the mismatch of the effective atomic number (Z(eff)) with the soil (mainly SiO2), good agreement was observed between the doses measured by TLDs and OSL dosimeters or the assessed doses. LiF:Mg,Cu,Si TLD has been found to be an effective alternative for the measurement of environmental gamma-ray annual dose rates for use in optical dating. (c) 2013 Elsevier Ltd. All rights reserved.

키워드

LiF:Mg,Cu,Si dosimeterAnnual gamma dose rateOptical datingLUMINESCENCEQUARTZ
제목
Environmental gamma-ray dose rate measurement by using ultra-high sensitive LiF:Mg,Cu,Si TLD
저자
Kim, M. J.Kim, S. Y.Lee, J. I.Kim, J. L.Chang, I.Hong, D. G.
DOI
10.1016/j.radmeas.2013.02.011
발행일
2013-09
유형
Article; Proceedings Paper
저널명
Radiation Measurements
56