상세 보기
Optimal conditions of X-ray irradiation for accurate equivalent dose determination
- Yawata, T.;
- Hashimoto, T.;
- Takeuchi, T.;
- Hong, D. G.
WEB OF SCIENCE
6SCOPUS
6초록
Luminescence excited by a small X-ray generator (Varian VF-50J) instead of the usual Sr-90-Y-90 beta source was measured following optical stimulation. The optimal X-ray irradiation conditions have been assessed, including the irradiation dose rate and the thickness of Al-absorber, in the equivalent dose determination when a luminescence dating process is simulated with laboratory dosed quartz samples. The equivalent doses were found to greatly depend on both the thickness of Al-absorber and the X-ray dose rate. We suggest that the optimal X-ray irradiation conditions for accurate luminescence dating work can be obtained by applying the following: (i) use of an Al-absorber with a thickness of 200 mu m and (ii) X-ray dose rates of 0.38 Gy/min to 2.9 Gy/min. (c) 2007 Elsevier B.V. All rights reserved.
키워드
- 제목
- Optimal conditions of X-ray irradiation for accurate equivalent dose determination
- 저자
- Yawata, T.; Hashimoto, T.; Takeuchi, T.; Hong, D. G.
- 발행일
- 2007-05
- 유형
- Article
- 권
- 258
- 호
- 2
- 페이지
- 375 ~ 380