Propagation of Ultrasonic Guided Waves in an Acrylic Plate as a Cortical-bone-mimicking Phantom

Citations

WEB OF SCIENCE

5
Citations

SCOPUS

5

초록

The present study aims to investigate the propagation of ultrasonic guided waves in an acrylic plate as a cortical-bone-mimicking phantom. The velocities of the guided waves in a 5-mm-thick acrylic plate were measured by using the axial transmission technique. A pure A0 Lamb mode could be successfully launched in the 5-mm-thick acrylic plate through a time reversal process of Lamb waves, consistent with the fact that the time reversal process can automatically compensate for the dispersive nature of Lamb waves. The experimental velocities of the slow guided wave (SGW) and the time-reversed Lamb wave were found to be in reasonable agreement with the theoretical group velocity of the A0 Lamb mode, suggesting that both the SGW and the time-reversed Lamb wave excited in the 5-mm-thick acrylic plate correspond to the A0 Lamb mode. These results suggest that the time reversal process of Lamb waves can be usefully applied to noninvasive characterization of long cortical bones.

키워드

OsteoporosisCortical boneUltrasonic guided waveLamb waveGroup velocityTime reversal processLEAKY LAMB WAVESTIME-REVERSALBOVINE TIBIAFEASIBILITY
제목
Propagation of Ultrasonic Guided Waves in an Acrylic Plate as a Cortical-bone-mimicking Phantom
저자
Lee, Kang IlChoi, Bok Kyoung
DOI
10.3938/jkps.65.1858
발행일
2014-12
유형
Article
저널명
Journal of the Korean Physical Society
65
11
페이지
1858 ~ 1862