바이오용 압전디스크방식 마이크로 펌프 설계 및 제작(I) -링형 플랩밸브 모듈의 최적설계-

Design and Fabrication of PZT Disc Actuated Micro Pump for Bio-Applications (I): Optimal Design of Ring-shaped Flab Valve Module

초록

Though a micro pump is a crucial element in miniaturized bio-fluidic systems or drug delivery systems, it has some intrinsic disadvantages such as backward flow, pulsation of flow, low repeatability and producibility, and high cost. To overcome these limitation, a PZT disc actuated micro pump including a novel ring-shaped multi-flap check valve is introduced. To enhance the performance of the micro pump, the static behaviour of flap valve module and the micro fluidic behavior of fluids are numerically modelled and analyzed. As results of the numerical analysis, the optimal dimension and configuration of the ring-shaped multi flap valve are 1×0.05 mm and 2 (inlet side)×1 (outlet side) type. The obtained maximum flow rates and flow resolution are about 90 ml/min and 30 ㎕/stroke respectively.

키워드

MicrofluidicsMicro pumpPZT discRing-shaped metal flap valveBackward flow prevention마이크로 플루이딕스마이크로 펌프압전소자 디스크링형 금속플랩밸브역류방지
제목
바이오용 압전디스크방식 마이크로 펌프 설계 및 제작(I) -링형 플랩밸브 모듈의 최적설계-
제목 (타언어)
Design and Fabrication of PZT Disc Actuated Micro Pump for Bio-Applications (I): Optimal Design of Ring-shaped Flab Valve Module
저자
김형진서영호김병희
발행일
2012-06
유형
Y
저널명
한국생산제조학회지
21
3
페이지
355 ~ 361