Development and Performance Testing of a Time-resolved OSL Measurement System

초록

Background: Time-resolved optically stimulated luminescence (TR-OSL) is a very useful method for calculating the lifetimes of crystalline quartz and feldspar. Materials and Methods: A compact TR-OSL system was developed, comprising a heater assembly manufactured using Kanthal wire, 2 powerful blue light-emitting diodes (LED, LXHLPB02) for optical stimulation equipped with VIS liquid light guides, and a photomultiplier tube combined with an optical filter for luminescence detection. A pulse generated from the data acquisition board (NI PCI 6250) was used to initiate on/off signals in LED and TR-OSL measurements. Results and Discussion: The TR-OSL and background signals measured using this TR-OSL system using quartz samples were very similar to those reported in a previous study. Additionally, the lifetimes of the build-up and TR-OSL signals were calculated as 27.4 ± 2.2 μs and 30.3 ± 0.6 μs, respectively, in good agreement with the findings of a previous study. Conclusion: It was concluded that the developed TR-OSL system was very reliable for TR-OSL signal measurements and lifetime calculations.

키워드

Time-resolved OSLPulse stimulationLifetimeQuartz
제목
Development and Performance Testing of a Time-resolved OSL Measurement System
저자
홍덕균김명진
DOI
10.14407/jrpr.2017.42.1.69
발행일
2017-03
유형
Y
저널명
방사선방어학회지
42
1
페이지
69 ~ 76