Luminescence properties after X-ray irradiation for dosimetry

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

0

초록

To investigate the luminescence characteristics after exposure to X-ray radiation, we developed an independent, small X-ray irradiation system comprising a Varian VF-50J mini X-ray generator, a Pb collimator, a delay shutter, and an Al absorber. With this system, the apparent dose rate increased linearly to 0.8 Gy/s against the emission current for a 50 kV anode potential when the shutter was delayed for an initial 4 s and the Al absorber was 300 A mu m thick. In addition, an approximately 20 mm diameter sample area was irradiated homogeneously with X rays. Based on three-dimensional (3D) thermoluminescence (TL) spectra, the small X-ray irradiator was considered comparable to the conventional Sr-90/Y-90 beta source even though the TL intensity from beta irradiation was higher than that from X-ray irradiation. The single aliquot regenerative (SAR) growth curve for the small X-ray irradiator was identical to that for the beta source. Therefore, we concluded that the characteristics of the small X-ray irradiator and the conventional Sr-90/Y-90 beta source were similar and that X ray irradiation had the potential for being suitable for use in luminescence dosimetry.

키워드

Luminescence dosimetrySmall X-ray irradiator3D TL spectrumOSL growth curveALPHA-PARTICLESTHERMOLUMINESCENCEEFFICIENCYQUARTZ
제목
Luminescence properties after X-ray irradiation for dosimetry
저자
Hong, Duk-GeunKim, Myung-Jin
DOI
10.3938/jkps.68.1120
발행일
2016-05
유형
Article
저널명
Journal of the Korean Physical Society
68
9
페이지
1120 ~ 1125