초록

Epidural catheter migration is a well-known cause of failed anesthesia and complications. One of the factors that affect catheter movement is when patients change their position after skin fixation. We report a case of an epidural catheter placed without evidence of intravascular or subdural insertion that produced an insufficient block. A 36-year-old woman presented for ankle surgery under epidural anesthesia. Epidural block was conducted at the L3-4 intervertebral space with a catheter threaded 3 cm into the epidural space with the patient in a back flexion and lateral position. The total volume of injected anesthetic was 28 mL, including a 3-mL test dose. The final anesthesia level was L1. The planned operation was completed without a pneumatic tourniquet. A postoperative C-arm fluoroscopic image revealed that 1 side hole of the catheter had moved out of the epidural space. We think that a positional change after catheter fixation was the reason for catheter outmigration leading to insufficient analgesia, which was incompatible with the amount of local anesthetic injected. (C) 2016 Elsevier Inc. All rights reserved.

키워드

Catheter displacement; Epidural block; Epidural catheter; COMPUTED-TOMOGRAPHY; ANALGESIA; MIGRATION; FAILURE; LABOR; HYPOTHESIS; SPACE
제목
Partial displacement of epidural catheter after patient position change: A case report
저자
Shon, Yoon-Jung; Bae, Seung-Kil; Park, Jae-Woo; Kim, In-Nam; Huh, Jin
DOI
10.1016/j.jclinane.2016.10.013
발행일
2017-02
유형
Editorial Material
저널명
Journal of Clinical Anesthesia
권
37
페이지
17 ~ 20