Rapidly tracking the position of a radiation source using multiple radiation spectroscopy detectors with maximum likelihood position estimation

  • Kim, Hyundong
  • Lee, Seung-Jae
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

The location of a radiation source can be tracked by deriving the location from the number of measured radiation counts using multiple radiation spectroscopy detectors. However, this method we previously developed was found to produce errors as the area of radiation incident on the source and the radiation spectroscopy detector changed. In this study, a location tracking system using a radiation spectroscopy detector and maximum likelihood position estimation (MLPE) was developed to derive the exact location. The lookup table (LUT) of the MLPE was created through Geant4 Application of Tomographic Emission (GATE) simulation, and the experimental data were applied to the LUT created through simulation to estimate the location of the radiation source. As a result, the estimation was confirmed to be exceptionally accurate (97.07 +/- 2.96 %). By applying the LUT created through simulation to the experiment, the LUT creation time was shortened and the convenience was improved. In addition, the measurement time could be shortened by using a radiation spectroscopy detector with excellent sensitivity as well as the ability to determine the location with excellent accuracy.

키워드

Radiation spectroscopy detectorGATEMLPELUTRadiation detectionGATEPET
제목
Rapidly tracking the position of a radiation source using multiple radiation spectroscopy detectors with maximum likelihood position estimation
저자
Kim, HyundongLee, Seung-Jae
DOI
10.1016/j.net.2025.103671
발행일
2025-09
유형
Article
저널명
Nuclear Engineering and Technology
57
9
페이지
1 ~ 5