상세 보기
Can use of video laryngoscopes by emergency medical technicians facilitate endotracheal intubation during continuous chest compression? A manikin study
- Han, S. K.;
- Choi, P. C.;
- Hong, C. K.;
- Shin, D. H.;
- Na, J. U.;
- ... Cho, J. H.;
- 외 1명
WEB OF SCIENCE
0SCOPUS
0초록
Introduction: We conducted this study to evaluate the utility of two video laryngoscopes (VLs) [Pentax-AWS (AWS), GlideScope (GVL)], compared to the conventional Macintosh laryngoscope (ML), on endotracheal intubation (ETI) involving chest compressions by Level 1 Korean emergency medical technicians (EMTs) who are the equivalent of EMT-I in the United States. Methods: This was a randomised crossover simulation study. Fifty EMTs performed endotracheal intubation in randomised sequence following two different scenarios: normal airway and difficult airway. Results: In normal airway scenario, overall success rate did not differ between the three devices. However AWS required a shorter run-time (14.1 [10.9-19.8] seconds) to complete ETI (T-C) than ML (17.7 [13.5-21.3] seconds) (p=0.017). And both VLs showed a significant superiority over ML in time required to visualise vocal cords (T-VC, percentage of glottic opening (POGO) score, and incidence of dental compression (I-DC). In difficult airway scenario, overall success rate of both VLs was significantly higher than ML. The T-C of AWS (13.7 [11.2-16.9] seconds) and GVL (20.7 [15.1-25.9] seconds) was shorter than that of ML (24.7 [18.1-34.5] seconds) (p<0.001). The T-VC of GVL was significantly shorter than that of AWS and ML. The POGO score, I-DC, and ease of intubation were significantly superior with AWS, GVL, and ML, respectively. Conclusions: Video laryngoscopes can facilitate EMT performing a faster and easier intubation without interrupting chest compressions. Moreover, AWS improves the success rate comparing to ML in difficult airway management.
키워드
- 제목
- Can use of video laryngoscopes by emergency medical technicians facilitate endotracheal intubation during continuous chest compression? A manikin study
- 저자
- Han, S. K.; Choi, P. C.; Hong, C. K.; Shin, D. H.; Na, J. U.; Hwang, S. Y.; Cho, J. H.
- 발행일
- 2014-09
- 유형
- Article
- 권
- 21
- 호
- 5
- 페이지
- 308 ~ 315