Optimization of the design parameters for a thyroid care nuclide monitoring diverging collimator using Monte Carlo simulation

  • Han, Dong-Hee; 
  • Lee, Seung-Jae; 
  • Kim, Jang-Oh; 
  • Kwon, Da-Eun; 
  • Jung, Kyung-Hwan; 
  • ... Baek, Cheol-Ha
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

The purpose is to develop gamma cameras to monitor the environment using diverging collimators designed to respond to I-13(1) through Geant4 application for tomographic emission (GATE) tool. The detector system is 50.0 x 50.0 x 3 5 mm(3) and contains gadolinium aluminum gallium garnet crystals with a 62 x 62 array. Our results were obtained by classifying the parameters as general purpose, high sensitivity, and high resolution; we also evaluated the extent of improvement in performance compared to a pinhole collimator. The minimum collimator height to remove an in-image artifact was 45.0 mm, and the reference full width at half maximum (FWHM) was set to about 3.0 mm for general-purpose use; the general-purpose parameters were a collimator height of 50.0 mm, hole size of 1.0 mm, and septal thickness of 0.4 mm. The high-sensitivity parameters were 45.0, 1.2, and 0.4 mm; the high-resolution parameters were 50.0, 1.0, and 0 8 mm, and the FWHM was 3.06, 3.99, and 2.38 mm, respectively. Our comparison showed a relatively constant performance regardless of the point source location in the field of view.

키워드

I-13(1); Radiation monitoring; GATE; Diverging collimator; GAMMA-CAMERA
제목
Optimization of the design parameters for a thyroid care nuclide monitoring diverging collimator using Monte Carlo simulation
저자
Han, Dong-Hee; Lee, Seung-Jae; Kim, Jang-Oh; Kwon, Da-Eun; Jung, Kyung-Hwan; Baek, Cheol-Ha
DOI
10.1007/s40042-022-00569-w
발행일
2022-10
유형
Article
저널명
Journal of the Korean Physical Society
권
81
호
7
페이지
675 ~ 679