Development of defibrillation simulator with LC-Tank type inductance coupling position measurement system

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

In this study, we construct a paddle positioning system for defibrillation training simulators. The system consists of a chest manikin fabricated based on the standard somatotype of Koreans and simulator paddles similar to those of commercial defibrillators. Multiple coils are arranged orthogonally in both the chest manikin and paddles; their positioning is based on electromagnetic induction, such that the positions of the paddles can be successfully detected across the chest. The calculated and actual positions are compared using the measured signals. We find a mean error of 0.0891 mm for all positions, with a standard deviation of 0.6611 mm. In addition, we connect our paddle positioning system to a graphical user interface that enables visualization of the paddles' current locations.

키워드

SENSOR
제목
Development of defibrillation simulator with LC-Tank type inductance coupling position measurement system
저자
Lee, Young Seok; Oh, Su Hwan; Lee, Yong Suk; Her, Ae-Young; Chang, In Bae
DOI
10.1371/journal.pone.0214576
발행일
2019-03-29
유형
Article
저널명
PLoS One
권
14
호
3