Measurement of the speed of sound in trabecular bone by using a time reversal acoustics focusing system

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

A new method for measuring the speed of sound (SOS) in trabecular bone by using a time reversal acoustics (TRA) focusing system was proposed and validated with measurements obtained by using the conventional pulse-transmission technique. The SOS measured in 14 bovine femoral trabecular bone samples by using the two methods was highly correlated each other, although the SOS measured by using the TRA focusing system was slightly lower by an average of 2.2 m/s. The SOS measured by using the two methods showed high correlation coefficients of r = 0.92 with the apparent bone density, consistent with the behavior in human trabecular bone in vitro. These results prove the efficacy of the new method based on the principle of TRA to measure the SOS in trabecular bone.

키워드

OsteoporosisTrabecular boneApparent bone densityUltrasoundSpeed of soundTime reversal acousticsULTRASONIC CHARACTERIZATIONBACKSCATTER MEASUREMENTSIN-VITROTRANSMISSIONDENSITY
제목
Measurement of the speed of sound in trabecular bone by using a time reversal acoustics focusing system
저자
Lee, Kang IlChoi, Bok Kyoung
DOI
10.3938/jkps.65.1036
발행일
2014-10
유형
Article
저널명
Journal of the Korean Physical Society
65
7
페이지
1036 ~ 1039