Evaluation of measurement uncertainty for quantitative determination of chlorite and chlorate in fresh-cut vegetables using ion chromatography

  • 정성진; 
  • 김다솜; 
  • 이근영; 
  • 윤상순; 
  • 임호수; 
  • ... 김희갑; 
  • 외 1명
Citations

SCOPUS

3

초록

This study aimed to evaluate the measurement uncertainty for the quantitative determination of chlorite and chlorate in ready-to-eat fresh-cut vegetables using ion chromatography with a hydroxide-selective column. One gram of the homogenized sample in deionized water was sonicated and centrifuged at 8,500 rpm. The supernatant was purified by passing it through a Sep-Pak tC18 cartridge, followed by chromatographic determination using a Dionex IonPac AS27 column. The linearity of the calibration curves, recovery, repeatability, and reproducibility of the method were satisfactory. The method detection limit was estimated to be approximately 0.5 mg/kg. Each uncertainty component was evaluated separately, and the combined and expanded uncertainty values were calculated at the 95% confidence level. The measured concentrations for 3 mg/kg of chlorite and chlorate standard materials were 3.18±0.32 and 3.10±0.42 mg/kg, respectively. These results confirmed the reliability of the developed method for measuring the two chlorine-based oxyanions in freshcut vegetables.

키워드

chlorite; chlorate; fresh-cut vegetable; ion chromatography; measurement uncertainty
제목
Evaluation of measurement uncertainty for quantitative determination of chlorite and chlorate in fresh-cut vegetables using ion chromatography
저자
정성진; 김다솜; 이근영; 윤상순; 임호수; 정영림; 김희갑
DOI
10.9721/KJFST.2017.49.6.591
발행일
2017-12
유형
Y
저널명
한국식품과학회지
권
49
호
6
페이지
591 ~ 598