상세 보기
초록
We report on the preconcentration-enhanced fast collection of myoglobin protein for the rapid detection of myocardial infarction. We use a one-dimensional micro/nanofluidic chip for electrokinetic preconcentration and demonstrate that the preconcentration factor of 1 ng/ml Alexa Fluor 488-labeled myoglobin is similar to 1000 within 200 s, where the protein had a weak negative charge, thereby making it hard to perform electrokinetic trapping for neutral-like proteins. The potential feasibility with new assay strategies for use in a rapid immunoassay screening test for myocardial infarction is discussed. (C) 2013 The Japan Society of Applied Physics
키워드
PRECONCENTRATION; BIOMARKERS; MEMBRANE; PROTEINS
- 제목
- Fast Myoglobin Detection Using Nanofluidic Electrokinetic Trapping Technique
- 저자
- Chun, DongWon; Kim, Sang Hui; Song, Hyungwan; Kwak, Seungmin; Kim, YooChan; Seok, HyunGwang; Lee, Sang-Myung; Lee, Jeong Hoon
- 발행일
- 2013-01
- 유형
- Article
- 권
- 6
- 호
- 1