Superconducting joints using multifilament MgB2 wires for MRI application

  • Patel, Dipak
  • Matsumoto, Akiyoshi
  • Kumakura, Hiroaki
  • Maeda, Minoru
  • Kim, Sun-Hu
  • ... Choi, Seyong
  • 외 4명
Citations

WEB OF SCIENCE

23
Citations

SCOPUS

24

초록

We present superconducting joints using 36 filaments magnesium diboride (MgB2) wires in continuation joint orientation and their characterisation results for the first time. One of the joints demonstrated the critical temperature (offset) of 34.4 K and the critical current (I-c) of 38.5 A at 10 K in self-field. Unconventional I-c behaviour of the joint was observed at 20 K in self-field when the I-c was measured in a high to a low field due to the effect of field trapping in the MgB2 bulk within the joint. Elemental mapping at the interface between the MgB2 bulk and the filament revealed oxidation, which was attributed to being responsible for reducing the I-c of the joint. This work will notably contribute towards MgB2 superconducting joint technology development for magnetic resonance imaging (MRI) application. (C) 2021 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.

키워드

Magnesium diborideMgB2 superconducting jointMultifilament MgB2 wireCritical currentMRI application
제목
Superconducting joints using multifilament MgB2 wires for MRI application
저자
Patel, DipakMatsumoto, AkiyoshiKumakura, HiroakiMaeda, MinoruKim, Sun-HuLiang, HaoYamauchi, YusukeChoi, SeyongKim, Jung HoHossain, Md Shahriar A.
DOI
10.1016/j.scriptamat.2021.114156
발행일
2021-11
유형
Article
저널명
Scripta Materialia
204