Comparison of conventional and automated breast volume ultrasound in the description and characterization of solid breast masses based on BI-RADS features

  • Kim, Hyunji
  • Cha, Joo Hee
  • Oh, Ha-Yeun
  • Kim, Hak Hee
  • Shin, Hee Jung
  • 외 1명
Citations

WEB OF SCIENCE

17
Citations

SCOPUS

23

초록

To compare the performance of radiologists in the use of conventional ultrasound (US) and automated breast volume ultrasound (ABVU) for the characterization of benign and malignant solid breast masses based on breast imaging and reporting data system (BI-RADS) criteria. Conventional US and ABVU images were obtained in 87 patients with 106 solid breast masses (52 cancers, 54 benign lesions). Three experienced radiologists who were blinded to all examination results independently characterized the lesions and reported a BI-RADS assessment category and a level of suspicion of malignancy. The results were analyzed by calculation of Cohen's kappa coefficient and by receiver operating characteristic (ROC) analysis. Assessment of the agreement of conventional US and ABVU indicated that the posterior echo feature was the most discordant feature of seven features (kappa = 0.371 +/- A 0.225) and that orientation had the greatest agreement (kappa A = 0.608 +/- A 0.210). The final assessment showed substantial agreement (kappa A = 0.773 +/- A 0.104). The areas under the ROC curves (Az) for conventional US and ABVU were not statistically significant for each reader, but the mean Az values of conventional US and ABVU by multi-reader multi-case analysis were significantly different (conventional US 0.991, ABVU 0.963; 95 % CI -0.0471 to -0.0097). The means for sensitivity, specificity, positive predictive value, and negative predictive value of conventional US and ABVU did not differ significantly. There was substantial inter-observer agreement in the final assessment of solid breast masses by conventional US and ABVU. ROC analysis comparing the performance of conventional US and ABVU indicated a marginally significant difference in mean Az, but not in mean sensitivity, specificity, positive predictive value, or negative predictive value.

키워드

Breast neoplasms diagnosisBreast neoplasms USUltrasound (US)Comparative studiesAutomated breast volume ultrasoundMAMMOGRAPHYSONOGRAPHYLESIONSBENIGN
제목
Comparison of conventional and automated breast volume ultrasound in the description and characterization of solid breast masses based on BI-RADS features
저자
Kim, HyunjiCha, Joo HeeOh, Ha-YeunKim, Hak HeeShin, Hee JungChae, Eun Young
DOI
10.1007/s12282-012-0419-1
발행일
2014-07
유형
Article
저널명
Breast Cancer
21
4
페이지
423 ~ 428