Research on Change Factors for Accurate Autocontouring of Organs when RTPs

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

When planning treatment using Radiation Treatment Planning system (RTPs), errors in the contouring process, which is the pre-planning stage, reduce the accuracy of the treatment. Therefore, we wanted to evaluate the error rate of RTPs, Window Width (WW), Window Level (WL), Hounsfield Unit (HU) values at the set point, and the error rate by organ, which are the causes of errors in the contouring process. The results showed that the error rate of WW, WL and HU of certain areas decreased over time. There were no significant changes in the other variables. Therefore, when performing auto-contouring for each organ, appropriate WW, WL and HU values should be set and additional manual contouring should be performed to ensure the accuracy of the treatment plan.

키워드

electro-magnetic radiation treatment planning; electro-magnetic radiation therapy; auto-contouring
제목
Research on Change Factors for Accurate Autocontouring of Organs when RTPs
저자
Kim, Jeong-Ho; Han, Man-Seok; Jeon, Min-Cheol; Yoo, Se-Jong
DOI
10.4283/JMAG.2023.28.4.432
발행일
2023-12
유형
Article
저널명
Journal of Magnetics
권
28
호
4
페이지
432 ~ 437