상세 보기
Optimization using Monte Carlo simulations of the thickness of a variable collimator for radiation monitoring
- Han, Dong-Hee;
- Won, Jong-Hun;
- Lee, Seung-Jae;
- Lee, HakJae;
- Baek, Cheol-Ha
WEB OF SCIENCE
4SCOPUS
4초록
In this study, we used GATE v8.0 to optimize the design parameters for the layer thickness and the shape of holes in variable collimators with new methods for moving detectors and acquiring magnified images in real time. The optimized thickness was a combination of five layers with thicknesses of 6.0, 2.0, 4.0, 2.0, and 6.0 mm. Image magnification on variable and conventional pinhole collimators, designed with the same parameters, confirmed that the greater the magnification is, the more distinct the two sources are. The increase in the penetration and the scatter ratio caused by empty areas created during the process of implementing the pinhole form into layers resulted in an average sensitivity error of 3.95%, which reduced spatial resolution slightly.
키워드
- 제목
- Optimization using Monte Carlo simulations of the thickness of a variable collimator for radiation monitoring
- 저자
- Han, Dong-Hee; Won, Jong-Hun; Lee, Seung-Jae; Lee, HakJae; Baek, Cheol-Ha
- 발행일
- 2021-04
- 유형
- Article
- 권
- 78
- 호
- 7
- 페이지
- 627 ~ 633
- 언어
- ENG
- 출판사
- KOREAN PHYSICAL SOC
- 발행국가
- 대한민국
- 분량
- 7 페이지
- ISSN
- E 1976-8524
P 0374-4884