Determination of bevantolol in human plasma by high performance liquid chromatography using solid phase extraction technique

  • Trung, Tran Quoc
  • Long, Pham Hai
  • Al-Abd, Ahmed M.
  • Ku, Hyo Jeong
  • Lee, Ho Yoon
  • 외 2명
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

0

초록

A method was developed and fully validated for the determination of bevantolol, an adrenergic-receptor blocker, in human plasma. Bevantolol and betaxolol as internal standard (I.S) were extracted from 1 mL of human plasma by solid phase extraction technique using Sep-pak silica cartridge. Chromatographic separation was accomplished under isocratic conditions using a reverse-phase C-8 analytical column and mixture of dibasic ammonium phosphate (pH 5.7; 50mM)-acetonitrile (75:25, v/v) as mobile phase, with a detection wavelength at 220 nm. The method was proved to be specific by testing six different human plasma sources. Linearity was established for the concentration ranges of 40-1600 ng/mL with correlation coefficent of 0.9995. The lower limit of quantification 40 ng/mL with precision of 10.9% as C.V%.

키워드

bevantololbetaxololHPLCsolid phase extractionplasmaBETA-1-ADRENOCEPTOR ANTAGONISTHYDROCHLORIDE
제목
Determination of bevantolol in human plasma by high performance liquid chromatography using solid phase extraction technique
저자
Trung, Tran QuocLong, Pham HaiAl-Abd, Ahmed M.Ku, Hyo JeongLee, Ho YoonHwang, Sung JooKim, Kyeong Ho
DOI
10.1007/BF02978842
발행일
2007-07
유형
Article
저널명
Archives of Pharmacal Research
30
7
페이지
890 ~ 897