Analysis of Magnetic Resonance Signals from Diffusion Weighted Imaging using Compressed Sensitivity Encoding Technique

Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

The purpose of this study was to compare image quality between compressed sensitivity encoding (CS) and sensitivity encoding (SENSE) in single-shot turbo spin echo diffusion-weighted imaging (TSE-DWI). Signal-to-noise ratios (SNR), apparent diffusion coefficient (ADC) values, and geometric accuracies were measured with an American College of Radiology head phantom. CS showed a high mean SNR, less variation in ADC values, and reduced imaging acquisition time (21.7 %-39.3 %) compared with SENSE. In addition, there was no significant difference in geometric accuracy between SENSE and CS. In conclusion, in comparison with SENSE TSE-DWI, CS TSE-DWI resulted in a reduced imaging acquisition time and improved SNR, while maintaining image quality.

키워드

magnetic resonance imaging; magnetic resonance signal to noise ratio; sensitivity encoding; geometric accuracy; TO-NOISE RATIO; HEAD; MRI; DISTORTION; SENSE; NECK
제목
Analysis of Magnetic Resonance Signals from Diffusion Weighted Imaging using Compressed Sensitivity Encoding Technique
저자
Jang, Ji-Sung; Han, Yong-Soo; Jeong, Mi-Ae
DOI
10.4283/JMAG.2019.24.2.217
발행일
2019-06
유형
Article
저널명
Journal of Magnetics
권
24
호
2
페이지
217 ~ 222