This study quantitatively evaluated the source term of a linear accelerator according to target thickness for a 6-20 MeV electron beam using MCNP6. The elements of the target were tungsten and copper, and a composite target and single target were simulated by setting different thickness parameters depending on energy. The accumulation of energy generated through interaction with the collided target was evaluated at 0.1-mm intervals, and F6 tally was used. The results indicated that less than 3% reference error was maintained according to the MCNP recommendations. At 6, 8, 10, 15, 18, and 20 MeV, the energy accumulation peaks identified for each target were 0.3 mm in tungsten, 1.3 mm in copper, 1.5 mm in copper, 0.5 mm in tungsten, 0.5 mm in tungsten, and 0.5 mm in tungsten. For 8 and 10 MeV in a single target consisting only of copper, the movement of electrons was confirmed at the end of the target, and the proportion of escaped electrons was 0.00011% and 0.00181%, respectively.& COPY; 2023 Korean Nuclear Society, Published by Elsevier Korea LLC. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

키워드

Linear acceleratorSource termMCNP6EvaluationRADIOTHERAPY
제목
Evaluation by thickness of a linear accelerator target at 6-20 MeV electron beam in MCNP6
저자
Han, Dong-HeeJung, Kyung-HwanKim, Jang-OhKwon, Da-EunLee, Ki-YoonLee, Chang-Ho
DOI
10.1016/j.net.2023.02.026
발행일
2023-06
유형
Article
저널명
Nuclear Engineering and Technology
55
6
페이지
1994 ~ 1998