Design of a Depth of Interaction Detector Module for Use in a Magnetic Resonance Imaging System

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

In this study, we designed a depth of interaction (DOI) detector module for use in a magnetic resonance imaging (MRI) system. The detector uses a semiconductor photosensor that is unaffected by a magnetic field. The detector module consists of two layers of a crystal array that have different light sharing characteristics. To verify the performance of the detector module, simulations of the optical photon transport in the crystal array were performed with a DETECT2000 tool. Various combinations of the parameters, such as the reflector used and the treatment of crystal surfaces, were simulated and the characteristics were evaluated through a reconstructed flood image. In certain combinations of the simulation conditions, all of the crystal pixels could be identified along all directions. Thus, the proposed DOI detector module could extract the three-dimensional gamma ray interaction position from the two-dimensional flood image.

키워드

Depth of interaction; PET/MRI; surface treatment; DETECT2000
제목
Design of a Depth of Interaction Detector Module for Use in a Magnetic Resonance Imaging System
저자
Lee, Seung-Jae; Baek, Cheol-Ha
DOI
10.4283/JMAG.2019.24.4.725
발행일
2019-12
유형
Article
저널명
Journal of Magnetics
권
24
호
4
페이지
725 ~ 728