Volumetric tumor measurement using three-dimensional ultrasound:: In vitro phantom study on measurement accuracy under various scanning conditions

  • Park, SH
  • Choi, BI
  • Han, JK
  • Yoon, CJ
  • Lee, JW
  • ... Kim, SS
  • 외 1명
Citations

WEB OF SCIENCE

27
Citations

SCOPUS

19

초록

Factors influencing volumetric tumor measurement using 3-D ultrasound (US) were investigated. A 3-D US unit equipped with 4- to 8-MHz curved-array and 5- to 10-MHz linear-array mechanical volume transducers, and a US phantom made of 20 ham pieces (8.6 similar to 10.5 mL) embedded in agar gel that simulated hyperechoic tumors, were used. Volumetric tumor measurement was significantly affected by the position of US focus and tumor depth. When focus was at 2 cm and 8 cm below the transducer, 5-cm-deep tumors were measured 5.6%+/-2.8% (mean+/-SD) and 8.3%+/-2.2% bigger, respectively, than when focus was at 5 cm below the transducer, the same position as tumors (p<0.01). Tumors were measured 11.8%+/-2.9% bigger when they were 8-cm deep below the transducer than when they were 5-cm deep below the transducer (p<0.001). Setting focus at the same position as tumors and keeping tumor depth consistent during serial 3-D US examinations may be necessary for reliable volumetric assessment of tumors. (C) 2004 World Federation for Ultrasound in Medicine Biology.

키워드

ultrasoundthree-dimensionalvolumetrytumorphantomIn vitro studyULTRASONOGRAPHYCTUS
제목
Volumetric tumor measurement using three-dimensional ultrasound:: In vitro phantom study on measurement accuracy under various scanning conditions
저자
Park, SHChoi, BIHan, JKYoon, CJLee, JWKim, SSHan, H
DOI
10.1016/j.ultrasmedbio.2003.09.010
발행일
2004-01
유형
Article
저널명
Ultrasound in Medicine & Biology
30
1
페이지
27 ~ 34