초록

CdZnTe(CZT) is a promising material for X-ray and gamma-ray detectors. The metal contact on the detector could determine, to a large degree, the performance of the detector. We investigate the dependency of the detector performance on the metal-semiconductor contact at various temperatures. CZT resistive detectors (Pt/CZT/Pt and Au/CZT/Au structure) and a CZT Schottky detector (In/CZT/Au structure) were used in our work. The gold contact was de posited with an electroless deposition method, and the indium contact was made with an evaporation method. The I-V curve and the gamma-ray energy spectrum were measured for each detector at various detector temperatures. The detector temperature was varied from -5 degrees C to 50 degrees C. The CZT Schottky detector was seen to a more stable detector performance with changing temperatures.

키워드

CZTMetal-semiconductor contactSchottky detectorResistive detectorCDZNTE DETECTORSCDTE
제목
Effect of Temperature on the Performance of a CZT Radiation Detector
저자
Park, S. H.Ha, J. H.Lee, J. H.Kim, H. S.Cho, Y. H.Cheon, S. D.Hong, D. G.
DOI
10.3938/jkps.56.1079
발행일
2010-04
유형
Article
저널명
Journal of the Korean Physical Society
56
4
페이지
1079 ~ 1082