The optimum flow rate in high temperature capillary column ion chromatography

  • Pyo, D
  • Kim, H
Citations

WEB OF SCIENCE

10
Citations

SCOPUS

10

초록

The optimum flow rates for the nitrate ion were measured experimentally using 1 m x 50 mum i.d. Latex particle coated silica capillary. The Van Deemter curves for nitrate at 75, 110 and 150 degreesC were also obtained. Increasing the column temperature resulted in an increase of the optimum flow rate for the nitrate ion. The theoretically calculated optimum flow rates in 50 mum i.d. open tubular ion chromatography matches fairly well with actual experimental data. (C) 2001 Elsevier Science B.V. All rights reserved.

키워드

ion chromatographyhigh temperaturecapillary columnoptimum flow rateMICROCAPILLARY LIQUID-CHROMATOGRAPHYSEPARATIONSPHASELC
제목
The optimum flow rate in high temperature capillary column ion chromatography
저자
Pyo, DKim, H
DOI
10.1016/S0026-265X(01)00124-2
발행일
2001-12
유형
Article; Proceedings Paper
저널명
Microchemical Journal
70
3
페이지
159 ~ 163