Differential argentometric titration equipped with a reflectance detector for the determination of halogen anions

  • Hwang, Hoon
  • Eom, In Yong
Citations

SCOPUS

1

초록

A simple technique utilizing a stopped flow argentometric titration system equipped with a reflection detector for the analysis of aqueous binary mixtures of Br- and I- ions was analyzed. A quartz vessel containing 40.0 mL of the titrand (an aqueous mixture of Br- and I- ions) was placed in a dark compartment built to block incoming external light. The titrant (0.101 M AgNO<inf>3</inf>) when introduced from a 50-mL burette into the titrand through a solenoid valve, the precipitation reaction started to produce characteristic silver halide precipitates. The timer was programmed to switch the solenoid valve 'ON' only during the first two seconds (flow rate of 3.36 mL/min) in every one minute for the discrete additions of Ag+ ion into the titrand. The light from the radiation source was guided through glass tubing, which is shed on the outer surface of the vessel throughout the titration.

키워드

Argentometric titrationDetermination of halide anionsDifferential titrationReflectance detector
제목
Differential argentometric titration equipped with a reflectance detector for the determination of halogen anions
저자
Hwang, HoonEom, In Yong
DOI
10.5012/jkcs.2010.54.5.497
발행일
2010
유형
Article
저널명
대한화학회지
54
5
페이지
497 ~ 499