Preliminary results of a small X-ray irradiator for equivalent dose estimation

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

In most established method for luminescence dating, the ratio of Sr-90/Y-90 radionuclides has been employed to provide reference laboratory doses to sample. Recently a small X-ray irradiator (Varian VF-50J) was introduced as an alternative to the traditionally used radioactive sources. Test results are presented whether the use of the small X-ray irradiator is possible in the dating technique. Aspects of the assessment of the X-ray irradiator include consideration of the measured optically stimulated luminescence signal and the relationship between X-ray tube currents and dose rates, as well as the dependence of equivalent dose for various dose rates. The test performances are assessed in relation to use with heated quartz.

키워드

LUMINESCENCETHERMOLUMINESCENCE
제목
Preliminary results of a small X-ray irradiator for equivalent dose estimation
저자
Hong, D. G.Yawata, T.Hashimoto, T.
DOI
10.1007/s10967-007-6874-z
발행일
2007-08
유형
Article
저널명
Journal of Radioanalytical and Nuclear Chemistry
273
2
페이지
353 ~ 356