Changes in the Glow Curve of Thermoluminescent Dosimeter According to Magnetic Field Strength

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

The thermoluminescent dosimeter is the most commonly used instrument for measuring radiation dose. Based on the previous study that the glow curve changes in the magnetic field environment of a thermal luminescence dose device, the change of the glow curve according to the change of the magnetic field strength is evaluated. Accordingly, the change in the glow curve was analyzed by irradiating 0.5-2 cGy to the thermoluminescent chips with a radiation generator while changing the magnetic field strength using a permanent magnet. As a result of the evaluation, an average increase of 0.3% was observed at 90 similar to 120 degrees C, an average decrease of -10.6% at 120-160 degrees C, and an average increase of 3.2% at 160-260 degrees C. Therefore, when applying the thermoluminescent dosimeter, it is necessary to evaluate the dose considering the error rate according to the magnetic field strength.

키워드

thermal luminescence dosimeter; glow curve; magnetic field strength; exothermic temperature; dose evaluation; MONTE-CARLO; ENERGY-DEPENDENCE; TLD; SENSITIVITY; TLD-100; PHOTON; TLD-700; BEAM; FILM
제목
Changes in the Glow Curve of Thermoluminescent Dosimeter According to Magnetic Field Strength
저자
Kim, Jeong-Ho; Han, Man-Seok; Jeon, Min-Cheol
DOI
10.4283/JMAG.2022.27.4.468
발행일
2022-12
유형
Article
저널명
Journal of Magnetics
권
27
호
4