A noncontact intraocular pressure measurement device using a micro reflected air pressure sensor for the prediagnosis of glaucoma

Citations

WEB OF SCIENCE

10
Citations

SCOPUS

13

초록

This study investigates a novel, portable tonometer using a micro reflected air pressure sensor for the prediagnosis of glaucoma. Because glaucoma progresses slowly and is not painful, glaucoma patients require a portable prediagnosis system to periodically measure intraocular pressure at home. Conventionally, intraocular pressure is measured by an air-puff tonometer whereby the cornea is deformed by a short pulse of air pressure and the magnitude of the corneal deformation is measured by optic systems such as a combination of laser- and photodiodes. In this study, a micro reflected air pressure sensor was designed, fabricated, and tested in order to measure the magnitude of corneal deformation without optic systems. In an experimental study, artificial eyes with different internal pressures were fabricated and these pressures were measured by the aforementioned system.

키워드

APPLANATION TONOMETRYTONO-PENMODULUSEYE
제목
A noncontact intraocular pressure measurement device using a micro reflected air pressure sensor for the prediagnosis of glaucoma
저자
Kim, Kyoung HwanKim, Byeong HeeSeo, Young Ho
DOI
10.1088/0960-1317/22/3/035022
발행일
2012-03
유형
Article
저널명
Journal of Micromechanics and Microengineering
22
3