상세 보기
초록
An analytical CE method was developed for the enantiomeric purity determination of fluvastatin enantiomers. Fluvastatin enantiomers were separated on an uncoated fused silica with 100 mM-borate solution containing 30 mg/mL of (2-hydroxypropyl)-beta-cyclodextrin (HP-beta-CD) as running buffer and fenoprofen as an internal standard. The linearity was observed within a 400-700 mu g/mL concentration range (r(2) >= 0.995) for both fluvastatin enantiomers. The repeatability expressed as coefficient of variation (CV) of the method were 0.96 and 0.92% for (+)-3R, 5S and (-)- 3S, 5R-fluvastatin, respectively. The limit of detection and quantification for both fluvastatin enantiomers were 1.5 mu g/mL and 2.5 mu g/mL. respectively.
키워드
fluvastatin; fenoprofen; CE; chiral separation; enantiomer; PERFORMANCE LIQUID-CHROMATOGRAPHY; COA REDUCTASE INHIBITORS; HUMAN PLASMA; ELECTROCHROMATOGRAPHY/; PHARMACOKINETICS; VOLUNTEERS; ANION; ASSAY
- 제목
- Chiral separation of fluvastatin enantiomers by capillary electrophoresis
- 저자
- Trung, Tran Quoc; Dung, Phan Thanh; Hoan, Nguyen Ngoc; Kim, Dae Joong; Lee, Joo Huyn; Kim, Kyeong Ho
- 발행일
- 2008-08
- 유형
- Article
- 권
- 31
- 호
- 8
- 페이지
- 1066 ~ 1072