Quantification of clotiazepam in human plasma by gas chromatography-mass spectrometry
- Ahn, SH;
- Maeng, HJ;
- Koo, TS;
- Kim, DD;
- Shim, CK;
- 외 1명
WEB OF SCIENCE
5SCOPUS
5초록
An analytical procedure was developed and validated for the quantification of clotiazepam in human plasma. After subjecting plasma samples to solid-phase extraction, the extract was evaporated and the residue re-constituted. An aliquot of the mixture was injected onto a gas chromatography-mass spectrometry system. The detector response was linear for clotiazepam concentrations in the range of 5-200 ng/ml. Intraand inter-day precision for the assay over the concentration range was below 13.1 and 13.5%, and the accuracy ranged between 99.0-107.9% and 92.4-101.3%, respectively. The drug was found to be stable under various processing conditions used. The method is applicable to human pharmacokinetic studies of clotiazepam. (c) 2006 Elsevier B.V. All rights reserved.