Detector Design of PET Measuring Depth of Interaction through Energy Spectrum Analysis

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

A depth of interaction detector was designed using three layers of the scintillation pixel array and silicon photomultiplier through DETECT2000 simulation. The upper and lower layers used diffuse reflectors, and the middle layer used specular reflectors to distinguish the light signal arising from each layer. The signals obtained at each layer were characterized by different locations at which photoelectric peaks were formed in the energy spectrum, which could be analyzed to track the layers in which gamma-rays and scintillation pixels interacted. Based on the simulation results, the basic experiment was conducted to obtain the energy spectrum. In the experiment, three photoelectric peaks were identified in the energy spectrum as in the simulation, and the layers interacted with gamma-rays, and scintillation pixels for each layer were distinguished by analyzing each photoelectric peak. It is anticipated that by applying the detector design obtained from this study to future small animal PET operating with MRI, the depth at which gamma-rays and scintillators interact will be measurable, thus improving the spatial resolution outside the FOV.

키워드

depth of interaction; DETECT2000; Reflectors; PET; MRI; magnetic field
제목
Detector Design of PET Measuring Depth of Interaction through Energy Spectrum Analysis
저자
Lee, Seung-Jae; Baek, Cheol-Ha
DOI
10.4283/JMAG.2020.25.4.687
발행일
2020-12
유형
Article
저널명
Journal of Magnetics
권
25
호
4
페이지
687 ~ 690