Assessment of binocular alignment using the three-dimensional Strabismus Photo Analyzer

  • Yang, Hee Kyung
  • Han, Sang Beom
  • Hwang, Jeong-Min
  • Kim, Young Jae
  • Jeong, Chang Bu
  • 외 1명
Citations

WEB OF SCIENCE

28
Citations

SCOPUS

29

초록

Aim To evaluate the efficacy of the three-dimensional (3D) Strabismus Photo Analyzer for estimating binocular alignment using photographs. Methods Thirty-two subjects with exotropia, 30 subjects with esotropia and 38 orthotropic subjects were included. Two independent ophthalmologists examined the angle of deviation using the Krimsky test and prism and alternate cover test (PCT). Full-face photographs were obtained using an 8.2-megapixel digital single-lens reflex camera, and the images were analysed using the 3D Strabismus Photo Analyzer. The images were adjusted for age-dependent ophthalmic biometry and angle kappa. The main outcome measures were inter-observer variability, test-retest reliability and correlation between the angles of deviation measured by different methods. Results The 95% limit of agreement of inter-observer variability was +/- 3.5 degrees (6.1 prism dioptres (PD)), +/- 3.1 degrees (5.4 PD) and +/- 1.5 degrees (2.6 PD) for the Krimsky test, PCT and the 3D Strabismus Photo Analyzer, respectively. The test-retest reliability was +/- 2.8 degrees (4.9 PD) for the 3D Strabismus Photo Analyzer versus the Krimsky test. The results of the Krimsky test and 3D Strabismus Photo Analyzer showed a strong positive correlation. Conclusions The 3D Strabismus Photo Analyzer is a simple and reliable tool for measuring ocular deviation. It has excellent agreement with the Krimsky test and substantially improved reproducibility.

키워드

ANTERIOR-CHAMBER DEPTHHIRSCHBERG TESTAXIAL LENGTHCLINICAL CONDITIONSPHOTOGRAPHIC METHODCORNEAL DIAMETERANGLE KAPPAORBSCAN-IIEYECALIBRATION
제목
Assessment of binocular alignment using the three-dimensional Strabismus Photo Analyzer
저자
Yang, Hee KyungHan, Sang BeomHwang, Jeong-MinKim, Young JaeJeong, Chang BuKim, Kwang Gi
DOI
10.1136/bjophthalmol-2011-300305
발행일
2012-01
유형
Article
저널명
British Journal of Ophthalmology
96
1
페이지
78 ~ 82