Estimation of stroke volume based on air pressure in air tube with pneumatic pulsatile ventricular assist device

Citations

SCOPUS

1

초록

A ventricular assist device (VAD) is used for bridge to heart transplantation and heart diseases. Knowing the status of a pneumatic pulsatile VAD when implanting it into the body is important: when the velocity of blood flow through the VAD is slow, a thrombus may occur, and thrombosis can be fatal to a patient. In order to determine the state of a VAD, various sensors need to be implanted. Because this introduces the risk of infection and difficulties with sensor management, we developed a method for estimating the state of a VAD indirectly via the pressure in an air tube that can be measured in vitro. We compared the measured values to in vitro experimental results. The estimated and measured values showed some errors, but the accuracy can be improved by refining the estimation process to minimize the risk of infection. © 2014 The Korean Society of Mechanical Engineers.

키워드

Air PressureEstimation of Stroke VolumeVentricular Assist Device
제목
Estimation of stroke volume based on air pressure in air tube with pneumatic pulsatile ventricular assist device
저자
Kang, YuminLee, Jin-hongHer, KeunChoi, Seongwook
DOI
10.3795/KSME-B.2014.38.12.971
발행일
2014
유형
Article
저널명
대한기계학회논문집 B
38
12
페이지
971 ~ 974