Estimation of stroke volume based on air pressure in air tube with pneumatic pulsatile ventricular assist device
- Kang, Yumin;
- Lee, Jin-hong;
- Her, Keun;
- Choi, Seongwook
SCOPUS
1초록
A ventricular assist device (VAD) is used for bridge to heart transplantation and heart diseases. Knowing the status of a pneumatic pulsatile VAD when implanting it into the body is important: when the velocity of blood flow through the VAD is slow, a thrombus may occur, and thrombosis can be fatal to a patient. In order to determine the state of a VAD, various sensors need to be implanted. Because this introduces the risk of infection and difficulties with sensor management, we developed a method for estimating the state of a VAD indirectly via the pressure in an air tube that can be measured in vitro. We compared the measured values to in vitro experimental results. The estimated and measured values showed some errors, but the accuracy can be improved by refining the estimation process to minimize the risk of infection. © 2014 The Korean Society of Mechanical Engineers.