A metallic buildup cap for glass rod dosimeters in radiotherapeutic in vivo dosimetry

  • Jeon, Hosang; 
  • Lee, Ho Jeong; 
  • Kim, Dong Woon; 
  • Joo, Ji Hyeon; 
  • Baek, Cheol-Ha; 
  • 외 5명
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

1

초록

In vivo measurement of radiotherapeutic doses is an important procedure for ensuring the accuracy of radiation treatments. The glass rod dosimeter is a known dosimeter for in vivo dosimetry, which has high reproducibility and low dependencies with respect to major radiotherapeutic conditions. A hemispherical buildup cap made of brass was fabricated to avoid a large measurement error owing to the partial lack of buildup effect during the measurement on the patient's skin. The correction factors (CFs) were experimentally obtained and verified using a Monte Carlo simulation. A clinical trial was performed with 117 patients. As a result, the measured CF was consistent with the simulation by a 2.3% error, and the mean errors of each site were within 2%. The total mean error was 1.84%, with a standard deviation of 1.39%. To conclude, this methodology can provide high accuracy and convenience in radiotherapeutic in vivo measurements.

키워드

In vivo measurement; Quality assurance; Radiotherapy; Glass rod dosimeter
제목
A metallic buildup cap for glass rod dosimeters in radiotherapeutic in vivo dosimetry
저자
Jeon, Hosang; Lee, Ho Jeong; Kim, Dong Woon; Joo, Ji Hyeon; Baek, Cheol-Ha; Ki, Yongkan; Kim, Wontaek; Park, Dahl; Nam, Jiho; Kim, Dong Hyeon
DOI
10.1007/s40042-021-00105-2
발행일
2021-05
유형
Article
저널명
Journal of the Korean Physical Society
권
78
호
9
페이지
837 ~ 841