The use of a ventricular assist device (VAD) can raise the one-year survival rate without cardiac transplantation from 25% to 52%. However, malfunction of the VAD system causes 6% of VAD patients' deaths, which could possibly be avoided through the development of new VADs in which VAD malfunctions do not affect the patient's heart movement or hemodynamic state. A conventional VAD has an impeller or vane for propelling blood that can allow blood to regurgitate when the propelling force is weaker than the aortic pressure. In this paper, we developed a new pulsatile conduit-shaped VAD that has two valves. This device removes the possibility of blood regurgitation and has a small stationary area even when the pumping force is extremely weak. We estimated the characteristics of the device by measuring the outflow and the pressure of the pump in in-vitro and in-vivo experiments. © 2011 The Korean Society of Mechanical Engineers.
키워드
- 제목
- Blood flow and pressure evaluation for a pulsatile conduit-shaped ventricular assist device with structural characteristic of conduit shape
- 저자
- Kang, Seongmin; Choi, Seongwook
- 발행일
- 2011
- 유형
- Article
- 저널명
- 대한기계학회논문집 B
- 권
- 35
- 호
- 11
- 페이지
- 1191 ~ 1198