Design Concept of Hybrid Instrument for Laparoscopic Surgery and Its Verification Using Scale Model Test

Citations

WEB OF SCIENCE

18
Citations

SCOPUS

19

초록

This paper proposes a new design concept of hybrid instrument for single-port laparoscopic surgery (SPLS) and a new method of verification using a scaled-up prototype based on the principle of elastic similarity. The proposed concept is a hand-held instrument that uses a tendon-gear mechanism for dexterous movement of its end-effector and servomotors with flexible tendon-sheath transmission to maintain the dexterity by compensating the loss of output angle from tendon elongation during the manual operation. The kinematic relationship of the tendon-gear mechanism was derived mathematically, and the ratio of external moment to resistive flexural stiffness of the articulating joint was matched between the real-sized model and the large-scale prototype. Our scale model tests have shown good agreement between their input-output relationships under the equivalent loading conditions, and thus verified the validity of similarity analysis. Also, the proof-of-concept experiments have demonstrated the functionality of output loss compensation of the hybrid instrument. Our methodology can be used to simplify and speed up the prototype development process for SPLS by avoiding miniaturization challenges such as high precision manufacturing, which is costly and time-consuming.

키워드

Elastic similarityhybrid laparoscopic surgical instrumentminimally invasive surgery (MIS)scale model testsingle-port laparoscopic surgery (SPLS)tendon-gear mechanismSINGLE-SITE SURGERYPERFORMANCEFUTUREFORCE
제목
Design Concept of Hybrid Instrument for Laparoscopic Surgery and Its Verification Using Scale Model Test
저자
Kim, Yoon-HoPark, Yong-JaiIn, HyunKiJeong, Chang WookCho, Kyu-Jin
DOI
10.1109/TMECH.2015.2453321
발행일
2016-02
유형
Article
저널명
IEEE/ASME Transactions on Mechatronics
21
1
페이지
142 ~ 153